엣지엔도 엣지파일의 피로 파절 저항성과 절삭능력에 관한 저널, EdgeEndo, EdgeFile

J Endod. 2017 May;43(5):823-826. doi: 10.1016/j.joen.2016.12.026.


나이타이 로타리 파일의 회전피로에 온도가 미치는 영항

The Effect of Temperature on Cyclic Fatigue of Nickel-titanium Rotary Endodontic Instruments.

Dosanjh A1, Paurazas S2, Askar M2.

Author information

1

Department of Postgraduate Endodontics, University of Detroit Mercy, Detroit, Michigan. Electronic address: andydosanjh@gmail.com.

2

Department of Postgraduate Endodontics, University of Detroit Mercy, Detroit, Michigan.

Abstract

INTRODUCTION:

This study examined the effect of different temperatures on the cyclic fatigue of nickel-titanium rotary files.

각기 다른 온도에서 니켈티타늄 파일의 회전피로도에 온오가 미치는 영향

METHODS:

Three groups of nickel-titanium rotary files (EF group [EdgeFile; EdgeEndo, Albuquerque, NM], VB group [Vortex Blue; Dentsply Tulsa Dental Specialties, Tulsa, OK], and ESX group [ESX; Brasseler USA, Savannah, GA]) of size 25 with a .04 taper and 25-mm length were tested in a metal block that simulated a canal curvature of 60° and a 5-mm radius curvature. The block was submerged in a water bath filled with water at 3°C, 22°C, 37°C, and 60°C. At each temperature, 30 files from each group were rotated at 500 rpm in the block. The number of cycles to fracture (NCF) was calculated. Statistical analysis was completed using a 1-way analysis of variance with significance at P < .05.

NCF: 회전수에 따른 파절

RESULTS:

VB group showed a significant decrease in NCF as the temperature increased from 3°C to 60°C. The ESX group showed a significant decrease in NCF as the temperature increased from 3°C to 37°C. The EF group showed a significant increase in NCF from 3°C to 22°C and a significant decrease in NCF from 22°C to 37°C. For each temperature, the EF group showed higher NCF than the VB group, which showed higher NCF than the ESX group.

각각의 온도에서 에지엔도의 에지파일이 가장 높은 NCF(회전수에 따른 파절, 파절저항성)를 보였다.

CONCLUSIONS:

In this in vitro study, temperature was found to significantly affect the cyclic fatigue of nickel-titanium rotary files. At each tested temperature, NCF was the highest for the EF group followed by the VB group and lowest for the ESX group. Future cyclic fatigue studies should be conducted at body temperature.

근관형성중 발생하는 열(온도)는 니켈티타늄파일 파일의 회전피로도에 큰 영향을 미친다. 각 온도 조건에서 에지엔도의 에지파일이 가장 강한 저항성을 보여 주었다.

Copyright © 2017 American Association of Endodontists. Published by Elsevier Inc. All rights reserved.

KEYWORDS:

Cyclic fatigue; files; nickel-titanium; temperature

J Endod. 2018 Mar;44(3):489-495. doi: 10.1016/j.joen.2017.09.008. Epub 2017 Dec 19.


에지엔도 나이타이파일의 절삭능력

Micro-computed Tomographic Evaluation of the Shaping Ability of XP-endo Shaper, iRaCe, and EdgeFile Systems in Long Oval-shaped Canals.

Versiani MA1, Carvalho KKT2, Mazzi-Chaves JF2, Sousa-Neto MD2.

Author information

1

Department of Restorative Dentistry, Dental School of Ribeirão Preto, University of São Paulo, Ribeirão Preto, SP, Brazil. Electronic address: marcoversiani@yahoo.com.

2

Department of Restorative Dentistry, Dental School of Ribeirão Preto, University of São Paulo, Ribeirão Preto, SP, Brazil.

Abstract

INTRODUCTION:

This study evaluated the shaping ability of the XP-endo Shaper (FKG Dentaire SA, La Chaux-de-Fonds, Switzerland), iRaCe (FKG Dentaire SA), and EdgeFile (EdgeEndo, Albuquerque, NM) systems using micro-computed tomographic (micro-CT) technology.

METHODS:

Thirty long oval-shaped canals from mandibular incisors were matched anatomically using micro-CT scanning (SkyScan1174v2; Bruker-microCT, Kontich, Belgium) and distributed into 3 groups (n = 10) according to the canal preparation protocol (ie, XP-endo Shaper, iRaCe, and EdgeFile systems). Coregistered images, before and after preparation, were evaluated for morphometric measurements of the volume, surface area, structure model index (SMI), untouched walls, area, perimeter, roundness, and diameter. Data were statistically compared between groups using the 1-way analysis of variance post hoc Tukey test and within groups with the paired sample t test (α = 5%).

RESULTS:

Within groups, preparation significantly increased all tested parameters (P < .05). No statistical difference was observed in the mean percentage increase of the volume (52%) and surface area (10.8%-14.2%) or the mean percentage of the remaining unprepared canal walls between groups (8.17%-9.83%) (P > .05). The XP-endo Shaper significantly altered the overall geometry of the root canal to a more conical shape (SMI = 2.59) when compared with the other groups (P < .05). After preparation protocols, changes in area, perimeter, roundness, and minor and major diameters of the root canals in the 5 mm of the root apex showed no difference between groups (P > .05).

CONCLUSIONS:

The XP-endo Shaper, iRaCe, and EdgeFile systems showed a similar shaping ability. Despite the XP-endo Shaper had significantly altered the overall geometry of the root canal to a more conical shape, neither technique was capable of completely preparing the long oval-shaped canals of mandibular incisors.

Copyright © 2017 American Association of Endodontists. Published by Elsevier Inc. All rights reserved.

테스트한 파일 모두 비슷한 절삭능력을 보였다.

KEYWORDS:

Micro–computed tomography; nickel-titanium instruments; reciprocating motion; root canal preparation; single-file system