The 44 years female patient presented as a referral from her general dentist due to a carious exposure of tooth #31 (Mandibular Right Second Molar). Upon completion of an exam and diagnosis, a treatment plan for endodontic therapy was presented to the patient and she subsequently consented to treatment. The case presented radiographically to demonstrate severe curvature approaching 90 degrees. Based upon previous experience and familiarity of Edge Endo files, I felt confident in navigating this canal system. This was after appropriate patency obtained with handfiles (Edge Endo K-Files).
EdgeFile™ with Fire-Wire™ Cyclic Fatigue Testing Study
Introduction
Heat treatment is used to change the deformation and strength characteristics of metals and metal alloys. The purpose of this evaluation was to determine how EdgeEndo’s™ EdgeFile™ with “Fire-Wire™“ NiTi files compared to various competitor NiTi endodontic rotary files performance when tested for cyclic fatigue.
Materials and Methods
EdgeEndo™ EdgeFile™ 20/06 with “Fire-Wire™” rotary files were compared to Dentsply Tulsa Dental® 20/06 Vortex Blue®, Dentsply Tulsa Dental® 20/06 GT Series X®, Dentsply Tulsa Dental® F1 ProTaper®, Sybron® 20/06 K3®, Sybron® 20/06 K3 XF®, and Brasseler® 20/06 Sequence® Endodontic Files. This study was performed at U.S. Endodontics in Johnson City, TN.
Cyclical Fatigue
Thirty of each file listed above and on the chart in the results portion of this study was evaluated for cyclical fatigue.
Cyclical fatigue test was performed using a custom test system (Fig. 1) with each file rotated at 500 RPM while 3mm of the file tip was inserted in a test block with a 90° curvature. Time to failure (seconds) was measured for each instrument.